Manual sensor tuning ==================== In manual tuning, the channel response curve should be: - (1) a sinusoids - (2) as large PTP amplitude as possible - (3) with the 0 from the Y-axis crossing at 1/3 from the bottom The BIAS controls the amplitude and the shape to some degree, the OFFSET the Y-axis crossing and the GATE controls the shape to some degree. The GATE should rarely be modified. .. tab-set:: .. tab-item:: Good tuning .. image:: ../_static/tuning/good.png :align: right :class: img-with-border :width: 100% .. tab-item:: Bad BIAS .. image:: ../_static/tuning/bad_bias.png :align: right :class: img-with-border :width: 100% .. tab-item:: Bad OFFSET .. image:: ../_static/tuning/bad_offset.png :align: right :class: img-with-border :width: 100% .. tab-item:: Bad GATE .. image:: ../_static/tuning/bad_gate.png :align: right :class: img-with-border :width: 100%