MEG contraindications๐
The signal measured by an MEG system is extremely low (\(10e^{-15}\ T\)), the device is very restrictive with potential artifacts-inducing elements. None of the contraindications listed below can be harmful for the participant or for the device: an MEG is a fully passive device. The contraindications will only impact the signal quality. Thus, if the cohort of participant is limited/rare it is possible to perform the experiment with a reduced signal quality. Depending on the artifact, it might be possible to correct in preprocessing through MaxWell filtering and Spatiotemporal SSS (tSSS).
Tip
MRI compatibility is not equivalent to MEG compatibility. An MRI compatible object might still contain some amount of ferromagnetic materials which do not disqualify it for MRI certification but will induce artifacts on the MEG signal.
Questionnaires for participants๐
Fillable PDFs in english or french are available below.
Known contraindications๐
pacemaker, or any kind of cardiac stimulator
insulin pump, bladder neurostimulator, deep-brain neurostimulator, or any kind of implanted device
cardiac valves, both mechanical or animal
hearing aid, hearing implant
joint prosthesis (hip, knee, shoulder, ..)[1]
neurosurgical clip for brain aneurysm
orthodontic equipment, dental prosthesis, dental retainer[2]
sutures with metal wires or staples
non-removable piercings, ear rings, ..
tattoos[1]
permanent makeup and eye makeup[3]
recent hair color or any kind of hair extensions
dreadlocks (MEGโs helmet is quite narrow)
foreign object in the eye
shrapnel or metal object in the body
filter in the vena cava to prevent pulmonary embolism
metal workers or construction workers working with metal[5]
On top of those contraindications, itโs better if the participant: